50mm×50mm

AM Solar Simulator – Basic Series

AM1.5G/AM0

The simulator can operate in both AM1.5G and AM0 by replacing the AM filter.

Optical Performance IEC 60904-9 Actual Value
Spectral Match (300–1200 nm)*1 87.5–112.5 % A+ 90–108 %
Irradiance Uniformity ≦2 % A 1.0-1.5 %
Temporal Instability STI ≦0.25 % A+ 0.14-0.23 %
LTI ≦1 % A+ 0.2-0.6 %

Air Mass Filter List

▲Menu

AM-Dependent
Air Mass Filter AM1.5G AM0
Standard Irradiance (=1 SUN)
280–4000 nm
1000 W/m² 1348 W/m²
1 SUN Available Time
(Not guaranteed)
≧3000 hours ≧2000 hours
Common Specifications
Initial Max. Irradiance*2 ≧1450 W/m²
Irradiance Adjustment Range*3 Approx. 70–100 %
Effective Irradiated Area*4 50×50 mm
Working Distance*5 174 mm
Irradiation Direction Downward
Lamp Model 150 W Xenon Lamp
External Shutter Operation ST Contact ON / OFF (Open / Close)
RY Contact Input (ON / OFF: Open / Close)
TT Not possible
(Open / Close time is set by Timer)
Input voltage 100-120 VAC 50/60 Hz, 240 W (Max.)
200-240 VAC
Packaging Information
(Reference)
41 × 47 × 78 cm, 18 kg (Solar Simulator)
58 × 39 × 50 cm, 10 kg (Stand)

▲Menu

A Irradiation ON / OFF By Shutter
B How to open and close Shutter Main unit switch / External switch
C Irradiance Adjustment By adjusting Lamp Current

▲Menu

A Irradiation ON / OFF By Shutter

BE

How to open and close Shutter

Main unit switch / External switch
Timers can be used to set opening and closing times
C Irradiance Adjustment By adjusting Lamp Current

▲Menu

Automatic forced air cooling after lamp shut-off Fan runs for 20 minutes, then stops
Interlock (If the cover is not attached properly) Lamp cannot be turned on,
or will be forcibly turned off.
Overheating prevention mechanism Lamp is forcibly turned off
by thermostat
D Lamp Hour Meter Total lighting time (resettable)

▲Menu

Voltage Endurance Input-Output / F.G:
Output / F.G:
1000 VAC – 1 min.
500 VAC – 1 min.
Insulation Resistance Input-Output / F.G:
Output / F.G:
DC 500 V above 20 MΩ
DC 500 V above 20 MΩ

▲Menu

Temperature 20-30 ℃ (RT) * The temperature change should be within 5 ℃
Humidity 20-85 % RH * No condensation

▲Menu

▲Menu

Metal Filters to control irradiance 10, 20, 30, 40, 50, 60, 70, 75 %
Additional Filter Dimensions 50 × 50 × 2 mm (t)

▲Menu