300mm×300mm

AM Solar Simulator – Basic Series

AM1.5G/AM0

Optical Performance IEC 60904-9 Actual Value
Spectral Match (300–1200 nm)*1 87.5–112.5 % A+ 90–110 %
Irradiance Uniformity ≦2 % A 0.9-1.2 %
Temporal Instability STI ≦0.5 % A 0.1-0.4 %
LTI ≦1 % A+ 0.4-0.8 %

Air Mass Filter List

▲Menu

AM-Dependent
Air Mass Filter AM1.5G
Standard Irradiance (=1 SUN)
280–4000 nm
1000 W/m²
1 SUN Available Time
(Not guaranteed)
≧1500 hours
Common Specifications
Initial Max. Irradiance*2 ≧1300 W/m²
Irradiance Adjustment Range*3 Approx. 70–100 %
Effective Irradiated Area*4 300×300 mm
Working Distance*5 765 mm
Irradiation Direction Downward
Lamp Model 1600 W Xenon Lamp
External Shutter Operation Contact Input (ON / OFF: Open / Close)
Input voltage 200-240 VAC
50/60 Hz, 2600 W (Max.)
Packaging Information
(Reference)
95 × 95 × 178 cm, 158 kg (Light Source)
82 × 68 × 62 cm,  49 kg (Power Supply)

▲Menu

A Irradiation ON / OFF By Shutter
B How to open and close Shutter Main unit switch / External switch
C Irradiance Adjustment By adjusting Lamp Current

▲Menu

Automatic forced air cooling after lamp shut-off Fan runs for 20 minutes, then stops
Interlock (If the cover is not attached properly) Lamp cannot be turned on,
or will be forcibly turned off.
Overheating prevention mechanism Lamp is forcibly turned off
by thermostat
D Lamp Hour Meter Total lighting time (resettable)

▲Menu

Voltage Endurance Input-Output / F.G:
Output / F.G:
1000 VAC – 1 min.
500 VAC – 1 min.
Insulation Resistance Input-Output / F.G:
Output / F.G:
DC 500 V above 20 MΩ
DC 500 V above 20 MΩ

▲Menu

Temperature 20-30 ℃ (RT) * The temperature change should be within 5 ℃
Humidity 20-85 % RH * No condensation

▲Menu

▲Menu

▲Menu